DHEA TEST RESULTS UPDATE
Labcorp has changed its DHEA testing method. The important distinction is that Labcorp has changed the analytical platform and established a new set of reference ranges. The new DHEA by Ionify LC-MS/MS (Test #004389) replaces the previous DHEA test (#004100).
The change also affects the units of measurement:
- Old DHEA test: ng/dL
- New DHEA Ionify test: ng/mL
What to watch for when processing reports
As SBN transitions its programming and reference ranges, please watch for blue-flagged DHEA results.
If you see a blue DHEA result:
- Check the unit of measurement on the Labcorp report against the unit shown in the SBN report. Make sure they match.
- First, re-analyze the report.
- If the DHEA result is still blue, submit a ticket through the Help with Data Entry section of the Ticket System to alert SBN.
If needed, you may also convert the result to the appropriate unit of measurement, manually enter the corrected value, and release the report for analysis.
How to convert DHEA results
Use the following conversions when the Labcorp and SBN units do not match:
NG/mL to NG/dL
Multiply by 100.
Example:
10.50 ng/mL × 100 = 1,050 ng/dL
NG/dL to NG/mL
Divide by 100.
Example:
1,050 ng/dL ÷ 100 = 10.50 ng/mL